Ludlum Measurements, Inc.  Ludlum Model 133-7

  Gamma Radiation Detector

 

 Specifications

 

Ludlum Model 133-2 Gamma Radiation DetectorPART NUMBER: 47-1216

APPLICATION: gamma radiation survey, dose rate, area monitoring
COMPATIBLE INSTRUMENTS: general purpose survey meters, ratemeters, scalers and area monitors
RADIATION DETECTOR: energy-compensated GM, 30 mg/cm2 stainless steel (halogen quench)
BACKGROUND: 1 cpm or less (10 μR/hr background)
LINEAR RANGE WITHOUT DTC: 25 mR/hr–30 R/hr (250 μSv/h–300 mSv/h)
LINEAR RANGE WITH DTC: 25 mR/hr–100 R/hr (250 μSv/h–1 Sv/h)
SENSITIVITY (137Cs gamma): 4.2 cpm/mR/hr (42 cpm/μSv/h)
ENERGY RESPONSE: within 15% of true value (60 keV-3 MeV)
DEAD TIME: typically 50 microseconds
OPERATING VOLTAGE: 460 volts
SIZE: 0.9 x 4 in. (2.3 x 10.2 cm) (Dia x L)
WEIGHT: 0.20 lb (0.090 kg)

CE CERTIFIED: yes

 

 Options & Accessories

 

Model 133-7-1 Waterproof Version (Part Number 47-2453) - Waterproof versions have the same characteristics as non-waterproof models, other than they are waterproof, contain an O-ring, and do not have a “C” connector on the probe end (base).  A 100 ft (30 m) cable is hard-wired into the detector. They weigh slightly more and have a slightly larger diameter than the non-waterproof versions.

 

Ludlum Model 133-4 Gamma Radiation Detector

Ludlum Model 133-4 Gamma Radiation Detector

Ludlum Model 133-4 Gamma Radiation Detector

Ludlum Model 133-4 Gamma Radiation Detector

Ludlum Radiation Detector Clips

Model 133-2

0.1 mR/hr-1 R/hr

Model 133-4

1 mR/hr-10 R/hr

Model 133-6

4 mR/hr-40 R/hr

Model 133-8

150 mR/hr-1000 R/hr

Detector Clip

 

0.25 µCi Cs-137 Radiation Check Source

     

 

Check Source

     

 

 

 Technical Information

 

Data Sheet

Product Manual

CE Certificate Energy Response

 

 Ludlum Resources

 

About Ludlum

Product Catalogs & Price Lists

Standard & Extended Warranty

Software Downloads

Replacement Parts

Repair & Calibration

Training