Ludlum Measurements, Inc. Logo  Ludlum Model 177

   Radiation Alarm Ratemeter

 

Ludlum Model 177 Alarm Radiation RatemeterFeatures

  • 4 Decade Range

  • Linear Analog Scale Readout

  • Adjustable Alarms

  • 50 Hour Rechargeable Battery

  • Supports GM and Scintillation Type Radiation Detectors

  • Total Counting Range from 0 - 500,000 cpm

 

Introduction

 

The Ludlum Model 177 is a rugged benchtop radiation ratemeter with alarming capability. These units are very popular for frisking operations at stations where line power is available but can also be used in portable applications for up to 50 hours using the internal rechargeable battery. The 177 supports G-M and scintillation type radiation detectors and displays the results on a linear readout analog meter. The audio alarm is adjustable from 0 through 150% of full scale and is factory set to latching. Reconfiguration to non-latching can be accommodated via a minor internal change. Front panel controls include a 4 position rotary switch for selecting each detector decade, alarm set control, audio volume adjust, fast/slow response switch, power on/off switch and push buttons for count reset, battery test, high voltage test and alarm test. The back panel contains the power cord connection and a 9 pin type "D" plug that supports, rapid battery charging, 6 - 9.6 volts power output, alarm output, pulse output and a recorder output.

 

 

Specifications

 

PART NUMBER: 48-1632

INDICATED USE: Multi-purpose radiation monitoring; frisking
COMPATIBLE RADIATION DETECTORS: G-M, scintillation
METER FACE: 0 - 500 cpm, 0 - 1.5 kV, BAT TEST (see picture below)
MULTIPLIERS: X1, X10, X100, X1k
LINEARITY: Reading within plus or minus 10% of true value with detector connected
CONNECTOR: Series "C" (others available)
AUDIO: Built in unimorph speaker with volume control (greater than 60 dB at 2 feet full volume)
ALARM: Indicated by a red lamp and full volume audible tone
DATA OUT: 9 pin series "D" connector allowing for external connections of the following items: battery; unregulated supply; instrument common ground; alarm signal (current sink); pulse out; external background subtract; and recorder output.
RECORDER OUTPUT: Correlated to meter reading (adjustable to 1 volt at 1 mA)
CALIBRATION CONTROLS: Accessible from back of instrument (protective cover provided)
HIGH VOLTAGE: Adjustable from 200 - 1500 volts (can be read on meter)
THRESHOLD: Adjustable from 10 - 100 mV
RESPONSE: Toggle switch for FAST (4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading
RESET: Push-button to zero meter
POWER: 95 - 135 VAC (178 - 240 VAC available), 50 - 60 Hz single phase (less than 100 mA), 6 volt sealed lead acid rechargeable battery (built in)
BATTERY LIFE: Typically 50 hours (battery condition can be checked on meter)
BATTERY CHARGER: Battery is continuously trickle charged when instrument is connected to line power and turned on (optional external fast charger is available)
METER: 2.5" (6.4 cm) arc, 1 mA analog type
CONSTRUCTION: Cast and drawn aluminum with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 5" (12.7 cm) H X 8" (20.3 cm) W X 6" (15.2 cm) L
WEIGHT: 4.2 lbs (1.9 kg)

 

This Product is CE Certified

 

 

Compatible Radiation Detectors

Compatible radiation detectors are listed below and are not included with the survey meter price. An additional charge will be added for each detector ordered. 

 

For Alpha
Detection

For Beta
Detection

For Gamma
Detection

For Alpha Beta Gamma Detection

 

Model 43-1*
Model 43-2*
Model 43-5*
Model 43-65*
Model 43-90*

Model 43-92*
Model 44-7
Model 44-9
Model 44-88
Model 44-89
Model 44-94

 

Model 44-1*
Model 44-6
Model 44-7
Model 44-9
Model 44-21
Model 44-38

Model 44-40
Model 44-88
Model 44-89
Model 44-94
Model 44-116*

Model 44-142*

Model 44-172*

 

Model 44-2
Model 44-3
Model 44-6
Model 44-7
Model 44-9
Model 44-10
Model 44-17
Model 44-20
Model 44-21
Model 44-38

Model 44-62
Model 133-2
Model 133-4
Model 133-6
Model 133-7
Model 133-8

 

 

Model 44-7
Model 44-9
Model 44-88
Model 44-89
Model 44-94

 

*For Mylar window scintillators it is strongly suggested that the optional overload indicator be used on the Model 177 Radiation Alarm Ratemeter to better determine the presence of a light leak.

 

 

Meter Face

 

 202-096
0 - 500 cpm; 0 - 1.5 kV

 

 

Options

  • Check Source

  • Overload Indicator

  • Low Battery Light Indicator Option - Model L-4347-155

  • 4-20 mA Output (Non-Isolated) - Model L-5173-351

  • 4-20 mA Outout (Isolated) - Model L-5347-079

  • Shelf (or Wall) Bracket - Model L-7173-078-02

  • Alarm Set Cover - Model L-7328-021

 

Replacement Parts

 

Technical Information

 

Other Ludlum Products

 

Ludlum Information

 

For more information or to request a quotation, please contact us

by phone at: (865) 947-5123 or by email at: sales@deqtech.com.