Ludlum Measurements, Inc. Logo  Ludlum Model 215

  Alpha Radiation Frisker

 

Ludlum Model 215 Alpha Radiation FriskerFeatures

  • Dual Purpose Alpha Radiation Frisker and Hand Monitor Station

  • Makes a great glove box radiation monitor

  • Detector Operates Independent of its Stand

  • Simple to Use

  • No Batteries, Cables or Gas Required

 

Introduction

 

The Ludlum Model 215 is a unique dual-purpose alpha radiation contamination detection system which can be used as both a stationary detector and mobile frisker. When being used alongside a glove box, the user can utilize this innovative device as a stationary hand frisker then simply remove it from the charging stand for use as they would any other mobile alpha radiation frisker.
 
This convenient, integrated detector design requires no batteries, P-10 counting gas or cables giving the user complete freedom in performing their frisking duties. The large area, air proportional radiation detector has built-in electronics, display and capacitors that allow it to be operated up to 30 minutes before needing to be recharged. Fully discharged, the detector is ready for stationary operation within 5 minutes of being placed into the charger stand and ready for mobile frisking within 20 minutes. The charging stand interfaces to a PC to facilitate setting parameters and running high voltage plateaus via an optional application program.

 

 

Specifications

 

PART NUMBER: 48-3531

INDICATED USE: Alpha radiation frisking, hand monitor with  alarm
RADIATION DETECTOR: Alpha air proportional probe with integrated electronics
WINDOW AREA: Active - 139 cm²; Open - 110 cm²
WINDOW: 0.4 mg/cm² aluminized mylar with aluminum screen (79% open). Can be easily replaced by removal of six screws on back of probe
EFFICIENCY: (4pi): 10%- 239Pu
DETECTOR USE TIME: When disconnected from charger, 30 minutes max
DETECTOR RECHARGE TIME: 20 minutes
DETECTOR DISPLAY:
Red LED - Alpha pulse or continuous on alarm
Amber LED - Trouble
Green LED - OK
DETECTOR AUDIO: 85 dB at 12 inches; programmable to other sound levels
OPERATING VOLTAGE: Altitude sensitive, 1950 V (sea level) to 1775 V (7000')
COUNTER THRESHOLD SETTING: 4 mV
CALIBRATION DUE DATE: When exceeded, amber trouble light will illuminate (real time clock located in charging/ calibration stand)
POWER:
Detector: Internal capacitors provide power for 30 minutes (in non-alarming condition) when operating independent of the stand.
Stand: +15 V (+12 V minimum to +20 V maximum) at 100 mA maximum.
CONSTRUCTION: Conductive plastic body with internal sealed electronics
TEMPERATURE RANGE: -4° F to 122° F (-20° C to 50°C)
SIZE:
Detector: 9.42” H x 4.38” W x 1.46” D (23.9 x 11.13 x 3.7 cm).
Charging/Calibration Stand: 9.0” H x 5.6” W x 2.78” D (22.9 x 14.2 x 7 cm)
WEIGHT:
Detector: 1.09 lbs. (0.49 kg)
Stand: 0.93 lbs (0.42 kg)

 

This Product is CE Certified

 

 

Ludlum Model 215-20 Charging/Calibration Stand

 

Options

 

Model 215-20 Charging/Calibration Stand - Part Number: 48-3567 interfaces with a personal computer and uses infrared communications to set parameters and run automatic plateau routines. The stand’s input power is +15V(+12V minimum to +20V maximum) at 100mA maximum. Four charging/calibration stands can be interconnected to run from a single 0.5A wall mount power supply.

 

Model 215 Alpha Frisker Station (Model 215 and Model 215-20 Charging/Calibration Stand) - Part Number: 48-3695

 

Model 215 Calibration Software - Part Number: 1370-080 available to assist in changing parameters and determining operational settings.

 

Ludlum Model 215 Alpha Radiation Frisker SoftwareThe Model 215 Calibration Software package allows the user to calibrate a Model 215 instrument and provides user-friendly access to the Model 215 features, including a general status check, as well as the ability to view and change calibration parameters, run a plateau, and calculate the efficiency of a Model 215 instrument.
 
The software interfaces with the instrument using an RS232 serial connection.  Instrument status parameters are updated every two seconds, allowing for continuous monitoring of an instrument, as well as the connection between the instrument and the charging stand.  The auto-update feature can be started and stopped at the discretion of the user.  The software also contains the Model 215 Debugger program, which can be accessed through the Debug menu on the main screen.

 

 

Spare Parts List

  • Model 215 Main Board (Part Number 5110-099)

  • Model 215-20 Charger Board (Part Number 5110-142)    

  • Model 215-20 Controller Board (Part Number  5110-172)

  • Model 215 Hex Screen (Part Number 7110-114)

  • Model 215 Window with Mylar (Part Number 4110-112)

  • Model 215 Entire Probe Body without Electronics (Part Number 4110-111)

  • Model 215 Electronics without Probe (Part Number 4110-116)

  • Model 215 Electronics Gasket (Part Number  7110-170)

  • Model 215-20 DC Power Supply - single plug (Part Number 21-9912) 

  • Model 215 Power Daisy Chain Cable 8’ (Part Number 8110-187)

 

Technical Information

 

Other Ludlum Products

 

Ludlum Information

 

For more information or to request a quotation, please contact us

by phone at: (865) 947-5123 or by email at: sales@deqtech.com.