Introduction |

● Ready-to-Go Radiation Emergency Response Kit
● Measures
Alpha, Beta, Gamma Contamination
● Measures
Radiation Rate,
Convenient Detector Switching
● High-Sensitivity
Gamma Probe for Quickly Finding Sources
● Auto-Ranging
Digital Radiation Ratemeter/Scaler
● Convenient Detector Switching
The Ludlum
Model 2241-3RK Radiation Emergency Response Kit packages all the basic radiation
measurement tools you are likely to need to handle a radiological emergency into
a convenient carrying case to facilitate a quick response when required. The kit
includes Ludlum's digital, auto-ranging and versatile
Model 2241-3 Ratemeter/Scaler. Responders can either plug in the
Model 44-9 external alpha beta gamma radiation GM pancake detector for
contamination measurements, the
Model 44-2 high sensitivity gamma radiation
scintillator for locating the source of the high readings or the
Model 133-7 energy compensated GM detector for exposure rate
measurements up to 100 R/hr. A convenient rotary
switch located on the front panel permits the responder to select the
appropriate operating parameter setup for the detector in use. The carrying case
supplied with the kit is air and water tight with cutouts for each device
including a
137Cs radioactive check source to verify proper operation of the
equipment.
|
Specifications |
PART NUMBER: 48-3189
APPLICATION: radiation emergency response
RADIATION DETECTORS: see Model 44-9, Model 44-2, Model 133-7 for external detector specifications
CONTROL SWITCH: A 5 position rotating switch to allow for quick change between four detector setups
METER TYPE: digital ratemeter and scaler
DISPLAY:
4-digit LCD display with 0.5 in. (1.3 cm) high digits, separate
annunciators for display units, alert, alarm, low battery, detector
overload, counting overflow, and scaler counting
BACKLIGHT:
pushbutton to activate
RATEMETER:
can display in R/hr, Sv/h, cpm, or cps when selector switch is in RATE
position
CONNECTOR: series "C" type
(others
available)
DISPLAY RANGE:
auto ranging from 0.0 μR/hr–9999 R/hr; 0 cpm–999 kcpm
BACKLIGHT “ON” TIME:
5, 15, 30, 60, 90, 120, 180, or 240 s for backlight to stay “ON” when
activated
SCALER:
adjustable from 1–9999 seconds (in 1-second intervals)
SCALER COUNT TIME:
adjustable from 1–9999 seconds
AUDIO DIVIDE:
1, 10, 100, or 1000 events per click
HIGH VOLTAGE:
400–2500 volts
DISCRIMINATOR:
adjustable from 2–100 mV
SET MINIMUM DISPLAY:
allows lower limit of auto-ranging display to be fixed, for example,
display can be set to not show values less than 1 mR/hr
DEAD TIME:
adjustable from 0–9999 microseconds
CALIBRATION CONSTANT:
adjustable from 0.001-280 X 109 counts/display unit
DETECTOR SWITCH:
rotating 5-position switch
DISCRIMINATOR ADJUSTMENTS:
1
RESPONSE:
VARIABLE: Dependent on number of counts present. Typical times are FAST 4 to 25 s, SLOW 4 to 60 s (All times stated from 10% to 90% of final reading).
FIXED: The parameter is adjustable from 10 to 25 resulting in a FAST response from approximately 2 to 50 s. The SLOW response is approximately 10 to 250 s. For MDA-type measurements, the fixed response mode is recommended.
ALARMS
ALERT/ALARM: indicated by annunciation on display
and audible tone
SCALER ALARM: adjustable from 1–999999 counts
RS-232 DATA DUMP MODE:
enables or disables binary dump mode to RS-232 port, when enabled binary
data will be dumped every 2 seconds
RS-232 DETECTOR SETUP MODE:
allows for detector parameters to be input by RS-232 port
BAUD RATE:
150, 300, 600, 1200, 2400, 4800, 9600, or 19200 bps
TEMPERATURE RANGE:
-4 to 122 °F (-20 to 50 °C)
POWER:
2 "D" cell batteries housed in an externally accessible sealed
compartment
BATTERY LIFE:
typically 200 hours with alkaline batteries (low battery warning on LCD)
CE CERTIFIED: yes
|
Options & Accessories |
|
|
|
Bluetooth Option 13 ft Range Part Number 4408-254 |
Bluetooth Option 30 ft Range Part Number 4408-302 |
|
|
|
|
|
|
Technical Information |
|
Ludlum Resources |