Introduction |
● Pocket-Sized
Alpha, Beta, Gamma Radiation Survey Meter
● 0 to 200 mR/hr (0 to 210kcpm)
● Internal End Window GM Detector
● 3 Decade Analog Meter
● Audio and Visual Alarms
The Ludlum Model 2401-EWA is a general purpose, pocket-sized radiation detection meter that employs an end-window type GM detector. The radiation detector is conveniently packaged inside the instrument with a protective screen covering located along the middle of the top end of the instrument. The meter presents two scales: CPM for radiation contamination measurements and mR/hr for radiation exposure rate and includes an audio and visual alarm.
|
Specifications |
PART NUMBER: 48-2997
APPLICATION: alpha, beta, gamma radiation survey
RADIATION DETECTOR: 0.57 in. (1.4 cm) diameter end-window GM
RANGE:
0–200 mR/hr; 0–210,000 cpm
SENSITIVITY:
typically 1050 cpm/mR/hr (137Cs
gamma)
ENERGY RESPONSE: energy dependent
RESPONSE TIME: typically 11 seconds from 10% to 90% of final reading
METER DIAL: 0–2 mR/hr; 0–2.1 kcpm, BAT OK
MULTIPLIERS: x1, x10, and x100
AUDIO: click-per-event, may be turned off by moving selector to QUIET position
POWER: 9 volt battery
BATTERY LIFE: typically 250 hours with alkaline battery at normal background levels
CONSTRUCTION: drawn and cast aluminum, with membrane panel
SIZE: 1.8 x 3.3 x 5.3 in. (4.6 x 8.4 x 13.5 cm) (H x W x L)
WEIGHT: 0.9 lb (0.4 kg), including battery
CE CERTIFIED: yes
|
Options & Accessories |
|
|
|
|
![]() |
|
|
Belt Clip Part Number 4397-176 |
Handle Part Number 4397-165 |
Air/Watertight Case Part Number 2311119 |
Canvas Case Part Number 2310517 |
Wrist Strap Part Number 4397-182 |
|
|
||||
|
|
|
Technical Information |
| Data
Sheet |
Product
Manual |
CE
Certificate |
|
Ludlum Resources |