Ludlum Measurements, Inc.  Ludlum Model 375P-336

  Waste Radiation Monitor

 

 Introduction

 

Ludlum Model 375P-336 Waste Radiation Monitor●  Model 375 Based Waste Radiation Monitor

●  Two 168 in3 Indoor Plastic Scintillation Detectors

●  Excellent Surface Contamination Screening Tool

●  Ideal for Monitoring Personnel, Laundry, Medical Waste

●  Also Monitors Small Articles, Baggage and Packages

●  Programmable Alarms, Battery Backup, Networkable

●  Outdoor Version Available - Model 375P-336-1

 

The Model 375P-336 is a Digital Model 375 Controller coupled to two 168 in³ (2753 cm³) plastic scintillation radiation detectors. The detectors and Model 375 Controller are indoor units and are typically wall mounted. This simple and cost-effective solution offers a simple system that is easy to operate and maintain. The controller supplies local alarms, but can also be connected to external alarms or even put onto an Ethernet network if desired. It additionally has a 24-hour battery backup to keep the system operational in the event power is lost. See Model 375 for additional detail and available options.
 

 Medical Application

 

In the nuclear medicine department, the Model 375P-336 is ideal for monitoring personnel or laundry for possible contamination that may have been unknowingly spilled. The monitor may also be used as a radiation contamination triage device, to alert emergency department personnel of potentially contaminated patients or equipment coming into the emergency room.

 

 Surface Contamination Inspection

 

Inspecting items using hand held instruments can be too time consuming in many cases. The Model 375P-336 facilitates a more rapid and uniform inspection by placing two relatively large sized scintillation detectors in close proximity to the incoming or outgoing articles undergoing inspection. Both detectors are continuously monitored by the digital controller so any offending item can immediately trigger an alarm. Alarm conditions can be setup to automatically halt production conveyance devices, notify the central office, and even alert key personnel to initiate an immediate response.

 

 Specifications

 

Ludlum Model 375-336 Waste Radiation MonitorPART NUMBER: 48-3285

APPLICATION: waste radiation monitoring
RADIATION DETECTORS: 2 ea. 168 in3 (2753 cm3) plastic scintillation detectors.  Each detector is supplied with a 50 ft (15.2 m) coaxial cable.
DISPLAY: 4-digit LED display with 0.8 in. (2 cm) digits

DISPLAY UNITS: μR/hr, mR/hr, R/hr, μSv/h, mSv/h, Sv/h, cpm, cps, and others

RANGE: 0.0 - 9999 kcps
LINEARITY: reading within 10% of true value
RESPONSE: typically 3 seconds from 10% to 90% of final reading
STATUS INDICATORS:

Ludlum Model 375-336 Waste Radiation MonitorINSTRUMENT STATUS: green light, instrument functioning properly
LOW ALARM: yellow light, slow audible beep (1/sec) audible tone (can be set from 0.0-9999)
HIGH ALARM: red light, fast beep (4/sec) audible tone (can be set from 0.0-9999)
OVERLOAD: senses detector saturation (indicated by display reading “-OL-”)
OVERRANGE: indicates radiation field being measured has exceeded counting range of instrument

HIGH VOLTAGE: adjustable from 200-2500 volts
THRESHOLD: adjustable from 2-100 mV
DEAD TIME: adjustable to compensate for dead time of detector and electronics (can be read on display)
AUDIO: can very from approximately 68 dB to 100 dB through operation of the external rotary baffle and the internal voltage connection

DATA OUTPUT: nine-pin connector providing five-decade logarithmic output, RS-232 output, signal ground connection, FAIL, and ALARM signals (current sink), and direct connection to battery and ground
CALIBRATION CONTROLS: accessible from front of instrument (protective cover provided)
POWER: a 9 Vdc wall-mount adapter, handles any mains voltage in the world, supplied with four sets of prongs for almost any style wall receptacle
BATTERY LIFE: typically 48 hours in non-alarm condition; 12 hours in alarm condition
BATTERY CHARGER: battery is continuously trickle charged when instrument is connected to line power and turned on
CONSTRUCTION: wall mount aluminum housing with ivory powder coat paint

SIZE: electronics: 10.3 x 9.7 x 3.3 in. (26.2 x 24.6 x 8.4 cm) (H X W x D); detectors (ea.): 41 x 6.8 x 2.1 in. (104 x 17.1 x 5.4 cm) (H x W x L)
WEIGHT: electronics: 9.3 lbs (4.2 kg); detectors (ea.): 23.5 lbs ( 10.7 kg)

CE CERTIFIED: yes

 

 Options & Accessories

 

Webpage and Service Software (Part Number 1370-077) Radiation network software package that collects and displays radiation levels and alarm status from up to fifty Model 375 type instruments. Using this interface, radiation data may be placed onto the Ethernet network for real-time monitoring, email alerts, and saved into a database. Webpage Hardware Interface (Part Number 4558-105) Data Sheet Product Manual

 

Ludlum Model 375 Webpage & Service Software

Ethernet Software (Part Number 1370-055) Monitors and stores data from up to 50 ethernet-enabled Model 375s. Instrument readings and status are displayed on the main screen. Ethernet Hardware Interface (Part Number 4558-098) Data Sheet Product Manual

 

Ludlum Model 375 Ethernet Software

Logging Software (Part Number 4396-167) Displays readings from the Model 375 in real time and saves the data to a database file. The software will also display a red HIGH ALARM, red DET FAIL, orange LOW ALARM, and a green STATUS button indicating the current status of the instrument. Ethernet Hardware Interface (Part Number 4558-098) Data Sheet Product Manual

 

Ludlum Model 375 Logging Software

Model 271 LED Remote Display (Part Number 48-2475) LED status indicators for STATUS OK, HIGH ALARM and DET FAIL. Sonalert-type speaker. 20 ft cable (Extra Cable: Part Number 8558-139). Power supplied by Model 375. Data Sheet Product Manual CE Certificate

 

Ludlum Model 271 Area Radiation Monitor Remote Display

Model 272 Analog Remote Display (Part Number 48-2656) Analog meter dial: 0.1 mR/hr–10 R/hr logarithmic scale. Status indicators for OK, ALARM and FAIL. Sonalert-type speaker. 20 ft cable (Extra Cable: Part Number 8558-140). Power supplied by Model 375. Data Sheet Product Manual CE Certificate

 

Ludlum Model 272 Area Radiation Monitor Remote Display

Model 272D Digital Remote Display (Part Number 48-3575) Four-digit LED display: 000.0-9999 with selectable units. Status indicators for STATUS, LOW ALARM, HIGH ALARM, DETECTOR FAIL and LOW BAT. Sonalert-type speaker. 20 ft cable (Extra Cable: Part Number 8558-141). Power supplied by 120-240 Vac with 5.0 Vdc regulated output.

Data Sheet Product Manual CE Certificate

 

Ludlum Model 272D Area Radiation Monitor Remote Display

Environmental Enclosure (Part Number 4396-068) This option inserts the Model 375 into a sealed, fiberglass, NEMA 4 rated enclosure suitable for outdoor or industrial type environments where water spray is likely to be encountered. Optional Strobe (Part Number 4396-1059). Optional keyed weather-resistant padlock (Part Number 4396-068)

 

Ludlum Model 375 Environmental Enclosure

Litestak Assembly Signal Enclosure (Part Number 4396-1001) Remote indicator for the Model 375. Includes 10 ft cable. (Extra Cable: Part Number 8396-1003)

 

Ludlum Model 375 Litestak

Strobe (110 V - Part Number 4558-147; 220 V - Part Number 4558-148) 50 ft Cable, Mains Relay Box, output on 3 pin Hirose connector.

Horn (110 V - Part Number 4558-149; 220 V - Part Number 4558-150) 50 ft Cable, Mains Relay Box, output on 3 pin Hirose connector.

Strobe & Horn (110 V - Part Number 4558-145; 220 V - Part Number 4558-146) 50 ft Cable, Mains Relay Box, output on 3 pin Hirose connector.

 

Ludlum Model 375 Horn/Strobe

Printer (Part Number 4558-185) with Date/Time stamping capability. 4.2 x 7.1 x 3.5 in. (W x D x H) 1 lb. Includes 20 ft RS-232 printer cable, paper roll, ribbon cassette, AC Adapter and manual.

 

Ludlum Model 375 Printer

Industrial Camera (Part Number 2311120) with high speed IP video and 1280 x 1024 resolution. 2.9 x 3.5 x 1.7 in. (H x W x D) 9 oz. Low-light performance, user-selectable exposure window, low power usage, multi-zone motion detection and digital pan, tilt and zoom. Simple to install and support. Optional Indoor/Outdoor Camera Housing (Part Number 21-9068)

 

Ludlum Model 375 Camera

Ludlum Model 375 Camera Housing

Output Options

Form C, Single Pole, Hi Alarm via 3 pin Hirose connector (Part Number 4396-201)

Dual Form C, Single Pole, via dual 3 pin Hirose connectors (Part Number 4396-501)

Mains Relay, via 3 pin Hirose connector (Part Number 4396-202)

Isolated 4-20 mA output (Part Number 4396-258)

 

 

 Technical Information

 

Data Sheet

Product Manual

CE Certificate

 

 Ludlum Resources

 

About Ludlum

Product Catalogs & Price Lists

Standard & Extended Warranty

Software Downloads

Replacement Parts

Repair & Calibration

Training