Introduction |
●
Application: Neutron Radiation Survey
● Detector: 3H Proportional
● Detection Range: Thermal to approximately 12 MeV
● Sensitivity: 100 cpm/mrem/hr (241AmBe)
● Gamma Rejection: ≤ 10 cpm through 10 R/hr
The Model
42-31H is a 9 in. REM ball type neutron radiation detector designed to be used with
portable counting instruments. It includes a top mounted bracket that
facilitates mechanical connection directly to a portable instrument's case. It
is designed for detection of thermal and fast neutrons
(0.025 eV to
approximately 12 MeV).
The Model 42-31H detector consists of a 3He detector
0.63 diameter x 1 in. long (1.6 x
2.5 cm), surrounded by a cadmium-loaded polyethylene sphere, 9 in. (22.9 cm) in diameter. A study is available that shows that the 9-inch
cadmium-loaded sphere has a response similar to that of a 10-inch diameter rem-responding
sphere.
|
Specifications |
PART NUMBER: 4005-140
APPLICATION: neutron radiation survey
RADIATION DETECTOR: 2 Atm ³He tube LND 25185 or equivalent
Moderator: 9 in. (22.9 cm)
diameter cadmium-loaded polyethylene sphere
Detection Range: thermal to approximately 12 MeV
Energy Response: thermal
to 12 MeV, follows radiation protection guide curve for neutron
dose
Sensitivity: typically 100
cpm/mrem/hr (1 cpm/mS/h) (241AmBe fast neutrons)
Gamma Rejection: typically
10 cpm or less through 10 R/hr (0.1 Sv/hr) (137Cs)
Input Sensitivity: -2 mV
Compatible Instruments:
typically used with portable counting instruments (capable of achieving -2 mV
input sensitivity)
Operating Voltage:
approximately 1200 volts
Connector: series “C”
(others
available)
Temperature Range: -4 to
122 °F (-20 to 50 °C)
Size: 10.3 x 9.0 x 9.0 in.
(26 x 22.9 x 22.9 cm) (H x W x D), including brackets
Weight: 14.5 lb (6.6 kg)
|
Technical Information |
|
Data
Sheet |
Product
Manual |
CE
Certificate |
|