Specifications |
PART NUMBER: 47-1305
APPLICATION:
alpha/beta radiation sample counting
RADIATION DETECTOR: ZnS(Ag) adhered to plastic scintillation disc
WINDOW: 0.4 mg/cm²
SAMPLE HOLDER: anodized aluminum tray with sample ring to allow for 1 in.
(2.5 cm) or 2 in. (5.1 cm) diameter samples
SAMPLE SIZE (maximum): 2 x 0.4 in. (5.1 x 0.9 cm) (Dia
x L)
EFFICIENCY (4π): 37% - 239Pu;
32% - 230Th; 39% -
238U; 5% -
14C; 27% -
99Tc; 29% -
137Cs; 26% -
99Sr/90Y
BACKGROUND:
alpha ≤ 3 cpm, beta ≤ 80
cpm
COMPATIBLE INSTRUMENTS: Ludlum Model
3030E
PHOTOMULTIPLIER TUBE: 2 in. (5.1 cm)
OPERATING VOLTAGE: 500-1200 volts
CONNECTOR: series
“C”
(others
available)
SIZE: 9.3 x 4.5 x 9.3 in. (23.6 x 11.4 x 23.6 cm) (H x W x L)
WEIGHT: 4.1 lb (1.9 kg)
CE CERTIFIED: yes
|
Options & Accessories |
![]() |
|
|||
| Sample Planchets |
|
|
Technical Information |
|
Data
Sheet |
Product
Manual |
|
Ludlum Resources |