Ludlum Measurements, Inc. Logo  Ludlum Model 43-135

  Large Area Radiation Source 2 pi Calibration Detector

 

 

Introduction

Ludlum Model 43-135 Large Area Radiation Source Calibration Detector
 
The Ludlum Model 43-135 is a large area 2 π proportional radiation detector designed for calibrating large area alpha and beta radiation sources.  The windowless detector and gold plated source tray are sealed together via heavy duty clamps which permit operating at higher pressures up to 0.2 MPa.   This detector design is in use at the National Institute of Standards and Technology  (NIST) and has demonstrated superior beta response and counting statistics with uncertainties in the range of approximately 0.1%.

 

 

Specifications

 

PART NUMBER: 47-3466

 

PERFORMANCE


Alpha background: ~ 0.0558 cps
Beta background: ~ 27.4 cps
Measurement uncertainty ~ 0.1%
Detector homogeneity response: ± 0.2%

DETECTOR


Detector area: 9.8” x 7.9”  (25 cm x 20 cm)   
Operating voltage: < 5000 Vdc
Plateau length:  >200 volts with a slope of ~1%, typical plateau from 1200 to 2700 volts
Typical count rates: 10 to 2000 cps
Detector connector: Hermetically sealed SHV connector (Ceramaseal  part number 9263-06A or equivalent)
Detector assembly seal:  Viton O-Ring

PRESSURE REQUIREMENTS


Working pressure:  22.1 to 29.4 PSIA, 1.5 to 2 atmospheres absolute (152.4 - 202.8 kPa)
Maximum pressure: 45 PSIA  (310.3 kPa)
Equipped with:
 Counting gas valves for inlet and outlet
 Pressure relief valve for over pressure protection
 Pressure gauge, 30” to 30 PSI range (vacuum to 30 psig)

PHYSICAL


Size: 21” deep x 17” wide x 7.2: high  (53.3 cm x 43.2 cm x 18.3 cm)
Maximum source dimension: 8.5” x 5.5” 0.5” tall (21.6 cm x 14 cm x 1.3 cm)
Weight: 82 lbs (37.2 kg)

 

 