Ludlum Model
43-68
Alpha Beta Radiation
Gas Proportional Detector - 100 cm²
Introduction
The Ludlum Model 43-68 is a 100 cm² large area gas proportional
alpha beta radiation detector for use in simultaneous alpha beta radiation surveys as well as alpha
radiation or beta
radiation surveys.
Specifications
PART NUMBER: 47-2005
INDICATED
USE: Alpha beta radiation survey
RECOMMENDED COUNTING GAS: P-10 (10% methane, 90% argon)
GAS RECHARGE: Will operate on a static charge for over 15 hours with a
39" cable
WINDOW: Typically 0.8 mg/cm² aluminized mylar (other thickness
available)
WINDOW AREA:
Active - 126 cm²
Open - 100 cm²
EFFICIENCY (4pi geometry): Typically 20% - 239Pu; 30% -
99Tc; 15% - 14C; less than 1% - gamma
BACKGROUND:
Alpha - Less than 5 cpm (when operating on the alpha only plateau region)
Beta - Typically 400 cpm or less (10 microR/hr field)
COMPATIBLE INSTRUMENTS:
Model 12,
Model 16,
Model 18,
Model 2000,
Model 2200,
Model 2221,
Model 2224-1,
Model 2350-1,
Model 2360
OPERATING VOLTAGE:
Alpha - Typically 1000 - 1500 volts
Beta/Gamma - Typically 1600 - 1800 volts
COUNTER THRESHOLD SETTING: Typically 2 - 5 mV
GAS CONNECTORS: Double end quick disconnect
CONNECTOR: Series "C"
(others available)
CONSTRUCTION: Anodized aluminum housing
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 3.9"(10cm)H X 4.6"(11.7cm)W X 7.8"(19.8cm)L
WEIGHT: 2 lbs (0.9kg)
Replacement Parts
Technical Information
Other Ludlum Products
Ludlum Information
For more information or to request a quotation,
please contact us
by phone at: (865)
947-5123 or by email at:
sales@deqtech.com.
|