Ludlum Model
43-89
Alpha Beta Radiation Scintillation Detector -
100 cm²
Introduction
The Ludlum Model 43-89 is a 100 cm² dual phosphor alpha beta
radiation scintillation detector that is designed to be used for simultaneously counting alpha and
beta radiation contamination.
Specifications
PART NUMBER: 47-2430
INDICATED
USE: Alpha beta radiation survey
RADIATION SCINTILLATOR: ZnS(Ag) adhered to 0.010"
thick plastic
scintillation material
WINDOW: Typically 1.2 mg/cm² aluminized mylar
WINDOW AREA:
Active - 125 cm²
Open - 100 cm²
EFFICIENCY (4pi geometry): Typically 16% - 239Pu;
5% - 99Tc; 16% - 90S/90Y
BACKGROUND:
Alpha - Less than 3 cpm
Beta - Typically 300 cpm or less (10 microR/hr field)
NON-UNIFORMITY: Less than 10%
CROSS TALK:
Alpha to Beta - Less than 10%
Beta to Alpha - Less than 1%
COMPATIBLE INSTRUMENTS: General purpose radiation survey meters, ratemeters,
scalers
TUBE: 1.5"(3.8cm) diameter magnetically shielded photomultiplier
OPERATING VOLTAGE: Typically 500 - 1200 volts
DYNODE STRING RESISTANCE: 100 megohm
CONNECTOR: Series "C"
(others available)
CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 5.5"(13.9cm)H X 4"(10.2cm)W X 12.3"(33cm)L
WEIGHT: 1.5 lb (0.7kg)
This Product is CE Certified