Ludlum Model
43-92
Alpha Radiation Scintillation Detector -
100 cm²
Introduction
The
Ludlum Model 43-92 is an Alpha Radiation Scintillation Detector. With an active area of 100 cm², a slim,
low-profile body, and a low background count rate, this detector is good for surveys for alpha radiation contamination.
Specifications
PART NUMBER: 47-2555
INDICATED USE:
Alpha radiation contamination survey
SCINTILLATOR: ZnS(Ag)
WINDOW: 0.8 mg/cm² recommended for outdoor use
WINDOW AREA:
Active - 100 cm²
Open - 89 cm²
EFFICIENCY (4pi geometry): Typically 20% - 239Pu
NON-UNIFORMITY: Less than 10%
BACKGROUND: Alpha - 3 cpm or less
COMPATIBLE INSTRUMENTS: General purpose radiation survey meters, ratemeters, and scalers
TUBE: 1.125"(2.9cm) diameter magnetically shielded
photomultiplier
OPERATING VOLTAGE: Typically 500 - 1200 volts
DYNODE STRING RESISTANCE: 100 megohm
CONNECTOR: Series “C”
(others available)
CONSTRUCTION: Aluminum housing with beige powder-coat finish
TEMPERATURE RANGE: -4°F(-20°C) to 122°F(50°C)
May be certified to operate from -40°F(-40°C) to 150°F(65°C)
SIZE: 3.2"(8.1 cm) H X 3.5"(8.9 cm) W X 12.2"(31 cm) L
WEIGHT: 1 lb (0.5kg)
This Product is CE Certified
Replacement Parts
Technical Information
Other Ludlum Products
Ludlum Information
For more information or to request a quotation,
please contact us
by phone at: (865)
947-5123 or by email at:
sales@deqtech.com.
|