Options & Accessories |
PART NUMBER: 47-1540
APPLICATION: low-level, wide-energy gamma radiation survey
RADIATION DETECTOR: 2 x 2 in. (5.1 x 5.1 cm) thick sodium iodide (NaI)Tl scintillator
EFFICIENCY: 125I - 4.0%
ENERGY RANGE: 50 keV-3.0 MeV
ENERGY RESPONSE: energy dependent (see graph below)
SENSITIVITY: typically 900 cpm/µR/hr (137Cs
gamma)
BACKGROUND: 9750 cpm
COMPATIBLE INSTRUMENTS: general purpose radiation survey meters,
ratemeters and scalers
PHOTOMULTIPLIER TUBE: 2 in. (5.1 cm) diameter
OPERATING VOLTAGE: 500-1200 volts
CONNECTOR: Series "C"
(others
available)
SIZE: 2.6 x 11 in. (6.6 x 27.9 cm) diameter (Dia x L)
WEIGHT: 2.3 lb (1.04kg)
CE CERTIFIED: yes

|
Options & Accessories |
|
|
|
|
|
|
|
Sample Holder |
Sample Holder |
Detector Collimator |
Detector Collimator |
|
|
|
|||
|
Replacement Parts |
NaI Crystal - 2 x 2 in. (5.1 x 5.1 cm): Part Number 01-5128
O-Ring: Part Number 16-8289
|
Technical Information |
| CE
Certificate |
Energy Response |
|
Ludlum Resources |