Ludlum Measurements, Inc. Logo  Ludlum Model 44-183

  Gamma Radiation Shielded G-M Detector

 

 

Features

  • Energy-Compensated GM Radiation Detector

  • Detector Range: 1 mR/h - 10 R/h

  • 10:1 Directional Shielding

  • Rugged Construction

 

Ludlum Model 44-183 Gamma Radiation Shielded G-M DetectorIntroduction

 

The Ludlum Model 44-183 facilitates gamma radiation surveys in high-background environments and in areas where elevated radiation fields are directional.  The detector design incorporates lead shielding that results in a 10:1 reduction in sensitivity from gammas entering in any direction but the bottom face.

The dynamic range is greatest with instruments equipped with DTC (Dead Time Correction) resulting in measurements of 1 mR/hr - 10 R/hr.  The GM radiation detector is a halogen quenched, energy-compensated GM.

The front plate and handle are constructed of aluminum. All exterior surfaces are protected by a durable, splatter-finish, beige colored paint.   Reference markings on the detector housing indicate the internal position of the embedded GM tube.

 

 

Specifications

 

PART NUMBER: 47-3758

DETECTOR RANGE WITH DTC: 1 mR/hr - 10 R/hr
DETECTOR RANGE WITHOUT DTC: 1 mR/hr - 2 R/hr
RADIATION DETECTOR: GM, energy-compensated
ENERGY RESPONSE: ± 15% of true value
SENSITIVITY (137Cs): 100 cpm/mR/hr (0.17 cps/μSv/h)
FRONT-TO-BACK SHIELD RATIO (137Cs): 10:1
OPERATING VOLTAGE: recommended 550 V
INPUT SENSITIVITY: -30 mV ± 10 mV
DEAD TIME: typically 50 microseconds
CONSTRUCTION: painted aluminum housing
CONNECTOR: series “C” (others available)
TEMPERATURE RANGE: -4 to 122 °F (-20 to 50 °C)
SIZE: 4.2 x 4.8 x 5.0 in (10.7 x 12.2 x 12.7 cm)  (H x W x L)
WEIGHT: 7.3 lb (3.3 kg)

 

This Product is CE Certified

 

 

 

Technical Information

 

Other Ludlum Products

 

Ludlum Information

 

For more information or to request a quotation, please contact us

by phone at: (865) 947-5123 or by email at: sales@deqtech.com.