Introduction |
● Pressurized
Ion Chamber (117 PSI)
● 0-5 R/hr Range with µR Sensitivity
● Rate, Integrate and Peak Readings
● Operator
Friendly Digital Meter with Color LCD
● Data
Logging
● USB Connectivity
● Sunlight Readable Color Display
● Auto Zeroing and Ranging
● Rechargeable Batteries
● Free Firmware Updates through Internet
The Ludlum Model 9DP, pressurized ion chamber radiation survey meter, provides highly sensitive measurements of exposure and exposure rate. It can simultaneously display the exposure rate and integrated value or the highest rate seen by the instrument. The integrated value can be reset (if desired) using one of the four convenient front-panel mounted buttons. The stunning 256-color, bit-mapped display provides an optimized presentation of the data and is accompanied with icons informing the user of the active functions and instrument status. All logged data are written in csv format to a plugged-in, industry standard, USB thumb drive for convenient retrieval by a PC spreadsheet or database program. Alarms are manifested using color changes on the display and an acknowledgeable audio output. The Model 9DP is part of Ludlum’s new Dimension series of meters employing state-of-the-art technologies that deliver tremendous capability, user friendliness, and convenient PC connectivity. The built-in USB port facilitates password-protected access to parameter settings via direct connection to a USB keyboard thus foregoing any need to install PC application software or dealing with operating system compatibility issues. Ludlum also offers an optional Dimension PC Windows™ interface program that enables total control over the instrument and performs calibration.
|
Specifications |
PART NUMBER: 48-3742
RADIATION
DETECTED: gamma and X-rays above 25 keV; beta
above 1 MeV
OPERATING RANGES:
0-500 μR/h, 0-5 mR/h, 0-50 mR/h, 0-500 mR /h, 0-5
R/h
CHAMBER VOLUME:
14 in³ (230 cc) volume pressurized to 117
psi (8 atmospheres), See
Model 9DP-1 for low pressure (25 psi) version.
ACCURACY:
+/-10%
RESPONSE TIME: 5 s in lowest range, 2
s in all other ranges, from 10% to 90% of final
value
MEASUREMENT READOUTS:
simultaneous display of rate and integrated reading or highest rate
DATA LOGGING:
Data is stored to detachable USB thumb drive in CSV format for easy
retrieval by PC spreadsheet/database programs. Data points include
real-time clock generated date and time with rate, integrated reading,
and instrument status. Logging time intervals are set by PC interface
program or standard USB keyboard.
LCD DISPLAY: 3.5 in. diagonal, 240 H x 320
W pixels, TFT
active matrix, 262,000 colors, 220 cd/m²
USER CONTROLS:
4 pushbuttons:
instrument on/off, peak
rate/integrate mode, audio on/off, alarm acknowledge/meter
reset/clearing integrated dose or peak rate
AUTOMATIC FUNCTIONS: auto
ranging, auto zeroing, auto LCD backlighting
AUDIO OUTPUTS:
built-in unimorph speaker, > 60 dB at 2 ft (0.6 m) An optional
audio
jack can be installed for connecting to an external headset.
ALARMS:
two levels of radiation alarms, each are user
programmable
throughout
entire readout range and set through a PC interface program. Others include low battery, various detector failures.
TEMPERATURE RANGE:
-4 to 122 °F (-20 to 50 °C)
POWER:
8 rechargeable "AA" NiMH batteries, supplied with wall charger
BATTERY LIFE:
12 to 30 hours between charges, depending upon use of backlighting
USB INTERFACE:
single USB port, may be connected directly
to a USB keyboard to
facilitate password-protected parameter changes, accepts USB thumbdrive
for storing logged data, optional interface kit facilitates connection
to a PC for parameter editing and calibration
CONSTRUCTION:
durable molded plastic with internal metal support
SIZE:
8.6 x 4.6 x 9.6 in. (21.9 x 11.6 x 24.5 cm) (H x W x L)
WEIGHT:
3.3 lb (1.5 kg), including batteries
CE CERTIFIED: yes
|
Options & Accessories |
Dimension Interface Package (Part
Number 4293-763) The Dimension Configuration Manager
software enables calibration and setup of any of the Dimension product
line of instruments. This package includes the Dimension Configuration
Manager software and a special USB cable for communications between the
PC and instrument. Software
Manual![]()

|
Data Logging Only Option |
|
|
![]() |
![]() |
|
Part Number 4293-998 |
Audio Jack Output Part Number 4293-891 |
Check Source | Storm Case |
|
Technical Information |
|
Ludlum Resources |