Introduction |
● Low Pressure
Pressurized
Ion Chamber (25 PSI)
● 0-50 R/hr Range, Non-Hazmat
● Special Design for Measuring Pulsed Fields
●
Operator
Friendly Digital Meter with Color LCD
● Data
Logging
● USB Connectivity
● Sunlight Readable Color Display
● Auto Zeroing and Ranging
● Rechargeable Batteries
● Free Firmware Updates through Internet
The Ludlum Model 9DP-1 ion chamber radiation survey meter is specially designed for radiography work where pulsed fields are being measured. This instrument correctly integrates 50 nanosecond pulses (and wider) that other systems typically miss or measure inaccurately. This instrument measures both exposure and exposure rate and can simultaneously display the exposure rate, integrated value, and highest rate seen by the instrument. The integrated value can be reset (if desired) using one of the four convenient front-panel mounted buttons. The buttons also control instrument power, function selection, setting the speaker volume and acknowledging alarms. The detector chamber is only pressurized to 25 psi, thus avoiding all (USA) HAZMAT concerns for shipping and handling. The stunning 256-color, bit-mapped display provides an optimized presentation of the data and is accompanied with icons informing the user of the active functions and instrument status. Alarms are manifested using color changes on the display and an acknowledgeable audio output. The instrument is powered using NiMH type rechargeable batteries that deliver up to 30 hours operation between charges. Measurements can be logged to an industry standard, USB thumb drive plugged into the instrument USB port. Data are written in csv format for convenient retrieval by a PC spreadsheet or database program. The Model 9DP-1 is part of Ludlum’s new Dimension series of meters employing state-of-the-art technologies that deliver tremendous capability, user friendliness, and convenient PC connectivity. The built-in USB port facilitates password-protected access to parameter settings via direct connection to a USB keyboard thus foregoing any need to install PC application software or dealing with operating system compatibility issues. Ludlum also offers an optional Dimension PC Windows™ interface program that enables total control over the instrument and performs calibration.
|
Specifications |
PART NUMBER: 48-3899
RADIATION
DETECTED: gamma and X-rays above 25 keV; beta
above 1 MeV, correctly integrates pulsed fields with 50 nanosecond
pulse widths
OPERATING RANGES:
0–5 mR/h, 0–50 mR/h, 0–500 mR /h, 0–5 R/h, 0–50
R/h
CHAMBER VOLUME:
13.4 in³ (220 cc) volume pressurized to 25 psi (1.7 atmospheres)
ACCURACY:
+/-10%
RESPONSE TIME:
5 s in lowest range, 2 s in all other ranges,
from 10% to 90% of final value
MEASUREMENT READOUTS:
simultaneous display of rate and integrated reading or
highest rate
INCLUDED
FUNCTIONS: integrated reading, peak
reading, range lock (0–50 R/h) for pulsed fields
DATA LOGGING:
Data is stored to detachable USB thumb drive in CSV format for easy
retrieval by PC spreadsheet/database programs. Data points include
real-time clock generated date and time with rate, integrated reading,
and instrument status. Logging time intervals are
set by PC interface
program or standard USB keyboard.
LCD DISPLAY:
3.5 in. diagonal, 240 H x 320 W pixels, TFT active matrix, 262,000
colors, 220 cd/m²
USER CONTROLS:
4 pushbuttons: instrument on/off, peak rate/integrate mode, audio
on/off, alarm acknowledge/meter reset/clearing integrated dose or peak
rate
AUTOMATIC FUNCTIONS:
auto ranging, auto zeroing, auto LCD backlighting
AUDIO OUTPUTS:
built-in unimorph speaker, > 60 dB at 2 ft (0.6 m) An optional
audio
jack can be installed for connecting to an external headset.
ALARMS:
two levels of radiation alarms, each are user programmable
throughout entire readout range and set through a PC interface program. Others include low battery, various detector failures.
TEMPERATURE RANGE:
-4 to 122 °F (-20 to 50 °C)
POWER:
8 rechargeable "AA" NiMH batteries, supplied with wall charger
BATTERY LIFE:
12 to 30 hours between charges, depending upon use of backlighting
USB INTERFACE:
single USB port, may be connected directly to a USB keyboard to
facilitate password-protected parameter changes, accepts USB thumbdrive
for storing logged data, optional interface kit facilitates connection
to a PC for parameter editing and calibration
CONSTRUCTION:
durable molded plastic with internal metal support
SIZE:
8.6 x 4.6 x 9.6 in. (21.9 x 11.6 x 24.5 cm) (H x W x L)
WEIGHT:
3.3 lb (1.5 kg), including batteries
CE CERTIFIED: yes
|
Options & Accessories |
Dimension Interface Package (Part
Number 4293-763) The Dimension Configuration Manager
software enables calibration and setup of any of the Dimension product
line of instruments. This package includes the Dimension Configuration
Manager software and a special USB cable for communications between the
PC and instrument. Software
Manual![]()

|
Data Logging Only Option |
|
|
![]() |
![]() |
|
Part Number 4293-998 |
Audio Jack Output Part Number 4293-891 |
Check Source | Storm Case |
|
Technical Information |
|
Ludlum Resources |