Ludlum Medical Physics  Ludlum Medical Physics Model L-051

  TG-51 Linac Filter

 

 

IntroductionLudlum Medical Physics TG-51 Linac Filter - Model L-051

 

The primary purpose of the AAPM TG-51 dosimetry protocol was to provide a uniform methodology for a clinical reference dosimetry measurement. Both the photon and electron beams from accelerators needed to be within the recommended nominal energies (Beam Quality). The methodology included the application of a 1mm thick lead foil that is placed just below the accelerator head, to reduce the electron contamination and therefore help to specify the beam quality. The lead foil is typically attached to the accelerator head or to the blocking tray using surgical tape, wires or whatever material was available.
 
The Ludlum Model L-051 “TG-51 Linac Filter” has been designed to simplify the task of making the prescribed Beam Quality Measurements, by providing a true 1mm thick lead foil (+/-0.2mm) that has been specially bonded to a 12” x12” polycarbon (Lexan ®) plate. The plate has an opening cut into the center, exposing a 10cm x10cm area of the bonded 11cm x 11cm lead foil. This plate can be customized by the user to fit the tracks of most blocking trays. The polycarbon material is easily cut with a standard utility knife or sheers. The combination lead foil and polycarbon plate also helps to maintain the integrity of the lead foil and also makes it easier to handle and store the filter.

 

 

Specifications

 

Part Number: 99-9432

12” x 12” Polycarbon Plate
11cm x 11cm x 1mm thick Lead foil

 

 