Ludlum Medical Physics Model L-600
Fluoroscopic Beam Alignment Tool
Introduction
The Ludlum
Fluoroscopic Beam Alignment device consists of an aluminum plate with 4 sliding
brass strips set in recessed channels. The strips define the border or visible
area of the image receptor. A plastic overlay prevents any vertical displacement
of the brass strips. Holes drilled in half inch intervals are filled with higher
density material for visibility through the brass strips. The device when placed
in the center of the image receptor is designed to correct or optimize
fluoroscopic collimation.
Any portion of the fluoroscopic field that falls outside the image receptor does
not contribute to the useful image and can lead to unnecessary exposure to the
patient. This very simple but critical measurement will identify a misaligned
fluoroscopic system.
Specifications
Part Number: 99-9406
Dimensions: 9”
x 9” x 5/8”
Weight: 5 Lbs.
Technical Information