Ludlum Medical Physics  Ludlum Medical Physics Model 2200

  Radiation Scaler/Ratemeter - Wipe Test Counter

 

 

FeaturesLudlum Medical Physics Scaler/Ratemeter - Wipe Test Counter - Model 2200

  • General Purpose Scaler/Ratemeter

  • Single Channel Analyzer

  • Supports Wide Range of External Detectors

  • 95-250 VAC Power Input

  • Portable Battery Operation

  • RS-232 Interface to Printer or PC

 

Introduction

 

The Ludlum Model 2200 Radiation Scaler/Ratemeter is the ideal economic solution for routine sample counting, single channel analyzing and routine radiopharmaceutical related procedures, when used with the Model 243 well scintillator (NaTl) detector. The Well Counter’s 4 pi geometry and excellent shielding provides excellent sensitivity to higher energy isotopes like I-131.

An adjustable discriminator and an adjustable window allow the user to count pulses within a user specified energy range. This portable unit can be powered by wall current or by four D-Cell batteries. The latter allows for continued operations during power interruptions. An optional printer (model 4167-386) is available for hardcopy archival of wipe test results.
 

 

Specifications

 

PART NUMBER: 48-1651

INDICATED USE: Single channel analyzing, gross counting
SUGGESTED DETECTORS: GM, proportional, scintillation
CONNECTOR: Series “C” (others available)
SCALER: 6 digit LED display providing a range of 0 - 999999 counts
SCALER LINEARITY: Reading within 2% of true value
TIMER: Pushwheel adjustment from 0 - 999 minutes with selectable X0.1 and X1 multipliers
RATEMETER: 0 - 500,000 cpm total range
METER DIAL: 0 - 500 cpm, 0 - 2.5 kV, BAT TEST
MULTIPLIERS: X1, X10, X100, X1000
RATEMETER LINEARITY: Reading within 10% of true value
RESPONSE: Toggle switch for FAST (4 seconds), or SLOW (22 seconds) from 10% to 90% of final reading
ZERO: Pushbutton to zero meter
HIGH VOLTAGE: Adjustable from 200 - 2500 volts (will support 60 megohm scintillation loads)
THRESHOLD: Adjustable from 1.00-10.00
WINDOW: Adjustable from 0 to 10.0 above the threshold setting (can be enabled or disabled)
DISCRIMINATOR: Adjustable from 2-100 mV at threshold setting of 1.00
RS-232: 9 pin connector allowing for printer or computer interface.
METER: 2.5" (6.4 cm) arc, 1 mA movement analog type
POWER: 85 - 250 Vac, 50-60 Hz or 4 each “D” cell batteries
BATTERY LIFE: Typically 120 hours with alkaline batteries (battery condition can be checked on meter)
TEMPERATURE RANGE: -4° F to 122° F (-20° C to 50° C) May be certified to operate from -40° F to 150° F (-40° C to 65° C)
SIZE: 8.5" H x 5" W x 9.3" D (21.6 x 12.7 x 23.5 cm)
WEIGHT: 6.9 lbs (3.1 kg) without batteries
 

Optional Printer Specifications


Can be setup to provide date/time stamp on each line.


POWER: comes with 120 Vac wall transformer, outputs 9 Vac
SPEED: 33 lines per minute
PAPER ROLL: standard 2.25" wide calculator paper
Ludlum Medical Physics Shielded Well Gamma Radiation Counter Head - Model 243
RIBBON LIFE: 200,000 characters
REPLACEMENT RIBBON: Epson HX-20 cartridge
SIZE: 4.05" W x 4.5" D x 2.0" H (10.3 x 11.4 x 5.1 cm)

 

 

Options

 

Technical Information

 

Other Ludlum Medical Physics Products

 

Ludlum Medical Physics Information

 

For more information or to request a quotation, please contact us

by phone at: (865) 947-5123 or by email at: sales@deqtech.com.