APPLICATION:
Gamma and X-ray exposure survey meter with beta response
RANGE:
0 - 50 R/hr
ENERGY RESPONSE:
+/- 20% of true value from 40 keV - 2 MeV
LINEARITY:
Reading within 10% of true value
RESPONSE TIME:
Approximately 5 seconds for 90% of final
meter deflection on the x1 and x10 scales, and 3 seconds on the
x100, x1k and x10k scales
BETA
RESPONSE: Factor of 4.8 difference
between window open and closed measurements when exposed to a
uranium slab
CHAMBER VOLUME: 220 cm³
CHAMBER
CONSTRUCTION: Carbon coated acrylic
SIDE WALL:
1000 mg/cm² aluminum and acrylic
BETA SHIELD:
Retractable 1000 mg/cm² phenolic slide
with side button control
WINDOW:
7 mg/cm² aluminized mylar
WINDOW AREA:
40 cm²
COMPENSATION:
Automatically corrects for temperature and pressure changes in
atmosphere
TEMPERATURE
RANGE: -4° to 122° F (-20° to 50°
C). Temperature compensation maintains calibration within 15% of
25° C reading
PRESSURE COMPENSATION
RANGE: 70 kPa to 106 kPa
METER:
2.5” (6.4 cm) arc, 1 mA, pivot-and jewel suspension
METER DIAL:
0 - 5 mR/hr, BAT TEST
CONTROLS:
-
Range Switch: 5 range multipliers x1,
x10, x100, x1k, x10k and instrument off
-
Reset: Pressing the reset button
causes the chamber to discharge
-
Bat Test: Pushbutton used to check
battery capacity
-
Zero Adjust: 1 turn potentiometer to
zero reading
-
Display Light: On/off switch
-
Calibration: Digitally set via USB to
serial computer interface, stored in non-volatile memory
BATTERY LIFE:
400 hours
POWER:
2 each “D” cell batteries housed in a
sealed externally-accessible compartment
CONSTRUCTION: Cast and drawn
aluminum with beige powder-coating
SIZE:
9.2” H x 3.5” W x 8.5” L (23.4 x 8.9 x 21.6 cm) including
instrument handle
WEIGHT:
4.2 lbs (1.9 kg) including batteries